1 Scout with a hand file #8 or #10 until it can
move smoothly along the canal
2 Determine working length via usual diagnostic
methods
3 PathFile #13 to working length, then
irrigate
4 PathFile #16 to working length, then
irrigate
5 PathFile #19 to working length, then
irrigate
6 Before starting the canal shaping with any
Nickel Titanium Rotary File System, confirm working length with a
K-File #15
7 Canal shaping with any Nickel Titanium Rotary
File System
Manufacturer’s recommendations:
Speed 300 Rpm
Torque 0.6 Ncm
|